39 Table 6. Pacemaker Interrogation and Programming Approach for CPP HBP LBBAP CRT Capture thresholds Determine His bundle capture relative to RV capture; program output to ensure His bundle capture (at least 1 V above the threshold) Determine LBB (LV septal) capture and anodal capture Determine LV (CS lead)–only capture and anodal capture Capture assessment algorithms Avoid, unless known that His bundle and RV capture thresholds are similar Capture assessment algorithms can be utilized successfully Capture assessment algorithms can be utilized successfully; LV–only pacing may be preferred in some cases AV delays Program 30–50 ms shorter than conventional parameters a Program 20–30 ms shorter than conventional parameters a Program 10–20 ms shorter than conventional parameters a Atrial oversensing Atrial oversensing can occur with proximal lead placement and may need appropriate programming to also avoid ventricular undersensing Ventricular unipolar sensing Avoid if pacing dependent a A shorter AV delay than conventional is needed to take account for the time delay from pacing output to QRS onset with conduction system pacing.