Page 30 of 47

31 His-Purkinje conduction disease With correction Without correction S-QRS ≤ wH-QRS with isoelectric interval S-QRS ≤ or > H-QRS with isoelectric interval Discrete local ventricular electrogram in HBP lead Discrete local ventricular electrogram in HBP lead Paced QRS < native QRS Paced QRS = native QRS 2 distinct capture thresholds (HBP with BBB correction, HBP without BBB correction) Single capture threshold (HBP with BBB) S-QRS < H-QRS (usually 0, S-QRS end < H-QRS end ) with or without isoelectric interval (pseudodelta wave +/−) S-QRS < H-QRS (usually 0) with or without isoelectric interval (pseudodelta wave +/−) Direct capture of local ventricular electrogram in HBP lead by stimulus artifact Direct capture of local ventricular electrogram in HBP lead by stimulus artifact Paced QRS ≤ native QRS Paced QRS > native QRS 3 distinct capture thresholds (HBP with BBB correction, HBP without BBB correction, RV capture) 2 distinct capture thresholds (HBP with BBB, RV capture)

Articles in this issue

Archives of this issue

view archives of HRS Guidelines Bundle (free trial) - Cardiac Physiologic Pacing for the Avoidance and Mitigation of Heart Failure