30 Implant Procedure Table 3. Criteria for His Bundle Pacing Baseline Normal QRS Selective HBP S-QRS = H-QRS with isoelectric interval Discrete local ventricular electrogram in HBP lead with S-V = H-V Paced QRS = native QRS Single capture threshold (His bundle) Nonselective HBP S-QRS < H-QRS (usually 0, S-QRS end = H-QRS end ) with or without isoelectric interval (pseudodelta wave +/−) Direct capture of local ventricular electrogram in HBP lead by stimulus artifact (local myocardial capture) Paced QRS > native QRS with normalization of precordial and limb lead axes with respect to rapid dV/dt components of the QRS 2 distinct capture thresholds (His bundle capture, RV capture) No QRS slur/notch in leads I, V1, or V4-V6, and V6 R-wave peak time ≤ 100 ms Change in V6 RWPT > 12 ms between stimulus and His to V6 RWPT confirms lack of His capture (99.1% sensitivity and 100% specificity) Adapted with permission from Vijayaraman, et al. Heart Rhythm 2018;15:460-468