Pump Evaluation Î Device parameters such as power, speed, flow, and pulsatility should be recorded throughout the patient's course. Table 4. Normal Parameters RPM Power, W Flow, L/min Pulsatility HeartMate II 8000–10 000 5.0–8.0 4.0–7.0 5.0–8.0 HeartWare Ventricular Assist Device 2400–3200 3.0–7.0 3.5–7.0 Î If battery power is low, either the batteries should be replaced or the configuration should be changed to wall power. Î When power source is confirmed but there is a persistent device alarm, the controller should be exchanged for the backup controller. Î If thrombus is suspected, assessment for hemolysis, including lactate dehydrogenase, is recommended.