27 Yellow alerts Device • Recommended replacement • Magnetic resonance imaging (MRI) mode Lead • Right atrial (RA)/RV/left ventricular (LV) ▶ Pacing impedance out of range ▶ Pacing threshold out of range Therapy • Single shock delivered • ATP delivered Pacing • RV pacing > programmed value • CRT pacing < programmed value Arrhythmias • AF burden > programmed value • Ventricular rate in atrial arrhythmia > programmed value • Non-sustained ventricular tachycardia (NSVT) in select patients Yellow alerts Device • Recommended replacement • MRI mode • In non-PM-dependent patients: ▶ Low battery voltage ▶ Device reset or in safety mode Lead • RA/RV/LV ▶ Pacing impedance out of range ▶ Pacing threshold out of range • Noise episode (non-PM dependent) Arrhythmias • AF burden > programmed value • Ventricular rate in atrial arrhythmia > programmed value • NSVT in select patients Pacing • RV pacing > programmed value • CRT pacing < programmed value Yellow alerts Battery repletion or reset Bradycardia • ≤30 bpm (without complete heart block) • Asystole/pause ≥ 3 seconds Arrhythmias • AF > 6 min with indication other than cryptogenic stroke Tachycardia • 16 beats > 180 bpm Red alerts are defined as critical alerts requiring urgent review. Yellow alerts are those that, with early review, may lead to an action that impacts patient outcomes.

Articles in this issue

Archives of this issue

view archives of HRS Guidelines Bundle (free trial) - Practical Management of the Remote Device Clinic