HRS Guidelines Bundle (free trial)

Practical Management of the Remote Device Clinic

Heart Rhythm Society GUIDELINES Apps brought to you free pf charge, courtesy of Guideline Central. Enjoy!

Issue link: https://eguideline.guidelinecentral.com/i/1499525

Contents of this Issue

Navigation

Page 13 of 39

14 Site-based RM COR LOE Recommendations 2a C-EO 1. For patients with CIEDs in centers without onsite device interrogation capability, it is reasonable to use site-based remote interrogation technolog y to facilitate access to care. 2a C-EO 2. For patients with CIEDs in centers with onsite device interrogation capability, it is reasonable to use site-based remote interrogation technolog y to provide expedited care. Technical Considerations in Remote Monitoring Devices with noncontinuous RM COR LOE Recommendations 1 C-EO 1. In patients with CIEDs on RM in the absence of continuous connectivity, remote transmissions are recommended at least every 3–12 months for PMs and every 3–6 months for ICDs. 1 C-EO 2. In patients with CIEDs on RM in the absence of continuous connectivity, as the device approaches elective replacement, the frequency of remote transmissions should be increased to every 1–3 months.

Articles in this issue

Archives of this issue

view archives of HRS Guidelines Bundle (free trial) - Practical Management of the Remote Device Clinic