ACCF/AHA Device-Based Therapy Guidelines

AHA GUIDELINES Apps brought to you courtesy of Guideline Central. All of these titles are available for purchase on our website, GuidelineCentral.com. Enjoy!

Issue link: https://eguideline.guidelinecentral.com/i/112380

Contents of this Issue

Navigation

Page 14 of 19

Table 1. Minimum Frequency of Cardiovascular Implantable Electronic Devices (CIEDs) In-Person or Remote Monitoringa Type and Frequency Method Pacemaker/ICD/CRT Within 72 h of CIED implantation In person 2-12 wk postimplantation In person Every 3-12 mo for pacemaker/CRTPacemaker In person or remote Every 3-6 mo for ICD/CRT-D In person or remote Annually until battery depletion In person Every 1-3 mo at signs of battery ��� depletion In person or remote Implantable loop recorder Every 1-6 mo depending on patient symptoms and indication In person or remote Implantable hemodynamic monitor Every 1-6 mo depending on indication In person or remote More frequent assessment as clinically indicated In person or remote a More frequent in-person or remote monitoring may be required for all the above devices as clinically indicated. Modified from Wilkoff B et al. Heart Rhythm 2008;5(6):907���25. 13